Performance Data Temperature
- Heating-up time to 150 °C [min]- 45
- Recovery time after 30 seconds door open at 150 °C [min]- 16
- Temperature fluctuation at 150 °C [± K]- 0.4
- Temperature range +5 °C above ambient temperature to: 300 °C
- Temperature variation at 150 °C [± K]- 1.5
Electrical data
- Nominal power [kW]- 1.35
- Phase (Nominal voltage)- 1~
- Power frequency [Hz]- 60
- Rated Voltage [V]- 120
- Unit fuse [A]- 12.5
Doors
- Unit doors-1
Housing dimensions not incl. fittings and connections
- Depth net [in]- 23.84
- Height net [in]- 27.78
- Width net [in]- 27.97
Internal Dimensions
- Depth [in]- 16.74
- Height [in]- 20.88
- Width [in]- 20.09
Measures
- Interior volume [cu.ft.]- 4
- Load per rack [lbs]- 66
- Net weight of the unit (empty) [lbs]- 126
- Permitted load [lbs]- 331
- Wall clearance back [in]- 6.3
- Wall clearance sidewise [in]- 3.94
Environment-specific data
- Energy consumption at 150 °C [Wh/h]- 245
Fixtures
- Number of shelves (std./max.)- 2/5