Performance Data Temperature
- Temperature range: -20...100 °C
- Temperature variation at 37 °C [± K]: 0.2 ± K
- Temperature fluctuation at 37 °C [± K]: 0.1 ± K
- Max. heat compensation at 37°C: 1,000 W
- Recovery time after door was opened for 30 s at 37°C: 4 mins
Electrical data
- Rated Voltage [V]: 208...240 V
- Power frequency [Hz]: 50/60
- Nominal power [kW]: 1.8 kW
- Unit fuse [A]: 16
- Phase (Nominal voltage): 1~
Measures
- Interior volume [cu.ft.]: 37.79
- Net weight of the unit (empty) [lbs]: 778.23
- Load per rack [lbs]: 88.18
- Permitted load [lbs]: 220.46
- Wall clearance back [in]: 3.94
- Wall clearance sidewise [in]: 7.09
External Dimensions (not including fittings or connections)
- Width [in]: 50.59
- Height [in]: 76.81
- Depth [in]: 31.43
Internal Dimensions
- Width [in]: 39.8
- Height [in]: 49.21
- Depth [in]: 33.34
- Inner Doors: 2
- Unit Doors: 2
Environment-specific data
- Sound-pressure level at 25°C: 50 dB(A)
- Energy consumption at 37 °C [Wh/h]: 340
Fixtures
- Number of shelves (std./max.): 2/16