- Operational environment: Indoor use only. Optimal operating range: +18°C to +26°C (+65°F to +78°F), 70% RH
- Exterior Dimensions (WxDxH): 23 5/8" x 23" x 20 3/8"
- Interior Dimensions (WxDxH): 17 3/4" x 16" x 12 7/8"
- Storage capacity: 1.7 cu. ft. gross volume
- Door: One swing solid door, self-closing, right hinged, non-reversible, magnetic sealed gasket, keyed lock
- Shelves: Three shelves (two adjustable/one fixed) with guard rail on back
- Mounting: Leveling legs. Note: 4" of clearance on all sides must be maintained for adequate ventilation
- Interior lighting: N/A
- External probe access: Rear wall port (3/8") dia.
- Insulation: Cabinet is foamed-in-place with EPA compliant high density urethane foam
- Exterior materials: White powder coated steel
- Access control: Pyxis®, Omnicell® and AcuDose RX® compatible
- General warranty: Two (2) years parts and labor warranty, excluding display probe calibration
- Compressor warranty: Five (5) years compressor warranty
- Product Weight: 80 lbs.
- Shipping Weight: 106 lbs.
- Rated Amperage: 1.0 Amps
- Power Plug/Power Cord: NEMA 5-15 plug, 8 to 10 ft typical, conforms to UL471 requirements, Vaccine storage power cord warning label
- Facility Electrical Requirement: 110-120V AC: 15 A (minimum)
- Agency Listing and Certification: Certified in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. UL, C-UL, ETL, C-ETL listed (either single or dual agency listings) and certified to UL471 standard, hydrocarbon refrigerant safety.
- Included Accessories:
- Pharmacy refrigerator/freezer toolkit and temperature logs
- Digital Data Logger (DDL) meets current CDC requirements for vaccine storage and monitoring. F/C switchable, has a 3-year certification of calibration, and (1) buffered probe in product simulated solution. Min/Max memory along with Alarm event handling. USB port for downloading stored data.
Performance
- Uniformity (Cabinet air): +/- 2.7°C
- Stability (Cabinet air): +/- 2.0°C
- Maximum temperature variation (Cabinet air): +/- 3.3°C
- Temperature rise after an after 8 sec door openings: Temperature did not exceed -19°C at any probe for all required NSF/ANSI 456 testing protocols
- Recovery after 3 min door opening: All probes recover to under -15°C within 3.2 min.
- Energy consumption: 0.75 KWh/day
- Average heat rejection: 1.27 KWh/day (180 BTU/h)
- Noise pressure level (dBA): 34 or less installed
- Pull down time to nominal operating temp: 100 min