- Operational environment: Indoor use only. Optimal operating range: +18°C to +26°C (+65°F to +78°F), 70% RH
- Exterior Dimensions (WxDxH): 23 3/4" x 24 1/2" x 33 3/8"
- Interior Dimensions (WxDxH): 19" x 17 1/4" x 21"
- Storage capacity: 4.2 cu. ft. gross volume
- Door: One swing solid door, self-closing, right hinged, non-reversible, magnetic sealed gasket, keyed lock
- Shelves: Two shelves, fixed
- Mounting: Low profile roller wheels and leveling legs
- Interior lighting: N/A
- External probe access: Rear wall port (3/4") dia.
- Insulation: Cabinet is foamed-in-place with EPA compliant high density urethane foam
- Exterior materials: White powder coated steel
- Access control: Pyxis®, Omnicell® and AcuDose RX® compatible
- General warranty: Two (2) years parts and labor warranty, excluding display probe calibration
- Compressor warranty: Five (5) years compressor warranty
- Product Weight: 100 lbs.
- Shipping Weight: 132 lbs.
- Rated Amperage: 1.5 Amps
- Power Plug/Power Cord: NEMA 5-15 plug, 8 to 10 ft typical, conforms to UL471 requirements, Vaccine storage power cord warning label
- Facility Electrical Requirement: 110-120V AC: 15 A (minimum)
- Agency Listing and Certification: Certified in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. UL, C-UL, ETL, C-ETL listed (either single or dual agency listings) and certified to UL471 standard, hydrocarbon refrigerant safety.
- Included Accessories:
- Pharmacy refrigerator/freezer toolkit and temperature logs
- Digital Data Logger (DDL) meets current CDC requirements for vaccine storage and monitoring. F/C switchable, has a 3-year certification of calibration, and (1) buffered probe in product simulated solution. Min/Max memory along with Alarm event handling. USB port for downloading stored data.
Performance
- Uniformity (Cabinet air): +/- 2.6°C
- Stability (Cabinet air): +/- 2.1°C
- Maximum temperature variation (Cabinet air): +/- 2.9°C
- Temperature rise after an after 8 sec door openings: Temperature did not exceed -17.5°C at any probe for all required NSF/ANSI 456 testing protocols
- Recovery after 3 min door opening: All probes recover to under -15°C within 8.2 min.
- Energy consumption: 0.97 KWh/day
- Average heat rejection: 1.97 KWh/day (224 BTU/h)
- Noise pressure level (dBA): 49 or less installed
- Pull down time to nominal operating temp: 51 min