- Operational environment: Indoor use only, +18°C to +26°C (+65°F to +78°F), <70% RH
- Storage capacity: 2.5 cu. ft. gross volume
- Exterior Dimensions (WxDxH): 17 3/4" x 20 7/8" x 30 3/4"
- Interior Dimensions (WxDxH): 14" x 13 1/4" x 21"
- Door: One swing solid door, self-closing, right hinged, non-reversible, magnetic sealed gasket, keyed lock
- Shelves: Four shelves (three adjustable/one fixed) with guard rail on back
- Mounting and Installation: Leveling legs
- Interior lighting: N/A
- Airflow management: Forced Air technology, patent pending
- External probe access: Rear wall port (3/8") dia.
- Insulation: Cabinet is foamed-in-place with EPA compliant high density urethane foam
- Exterior materials: White powder coated steel
- Access control: Pyxis, Omnicell and AcuDose RX compatible
- General warranty: Two (2) years parts and labor warranty, excluding display probe calibration
- Compressor warranty: Five (5) years compressor warranty
- Product Weight: 72 lbs.
- Shipping Weight: 88 lbs.
- Rated Amperage: 0.9 Amps
- Power Plug/Power Cord: NEMA 5-15 plug, 8 to 10 ft typical, conforms to UL471 requirements, Vaccine storage power cord warning label
- Facility Electrical Requirement: 110-120V AC: 15 A (minimum)
- Agency Listing and Certification: Certified with the temperature performance requirements as defined in the NSF/ANSI 456 Standard for Vaccine Storage for all testing scenarios. UL, C-UL, ETL, C-ETL listed and certified to UL471 standard, hydrocarbon refrigerant safety.
- Included Accessories:
- Pharmacy refrigerator/freezer toolkit and temperature logs
- Digital Data Logger (DDL) meets current CDC requirements for vaccine storage and monitoring. F/C switchable, has a 3-year certification of calibration, and (1) buffered probe in product simulated solution. Min/Max memory along with Alarm event handling. USB port for downloading stored data.
Performance
- Uniformity (Cabinet air): +/- 1.1°C
- Stability (Cabinet air): +/- 0.7°C
- Maximum temperature variation (Cabinet air): +/- 1.2°C
- Temperature rise after an after 8 sec door openings: Temperature did not exceed 7.2°C at any probe for all required NSF/ANSI 456 testing protocols
- Recovery after 3 min door opening: All probes recover to under 8°C within 7.5 min.
- Energy consumption: 0.58 KWh/day
- Average heat rejection: 1.0 KWh/day (142 BTU/h)
- Noise pressure level (dBA): 34 or less installed
- Pull down time to nominal operating temp: 42 min
Note: The CDC requires a data logger or temperature monitoring device to consistently record live temperature readings from the inside of a refrigerator or freezer when storing all types of vaccines. Constant monitoring of vaccines at recommended temperatures is essential for maintaining the efficacy of a vaccine. Failing to do so can lead to wasted vaccines or ineffective vaccines being administered to patients. Please follow the CDC requirements for Vaccine Storage & the COVID-19 Vaccination Program found here.
*Please note this unit ships freight and on a pallet.