- Operational environment: Indoor use only, +18°C to +26°C (+65°F to +78°F), <70% RH
- Storage capacity: 4.6 cu. ft. gross volume
- Exterior Dimensions (WxDxH): 23 7/8" x 24 3/8" x 33 3/8"
- Interior Dimensions (WxDxH): 19 1/4" x 17 1/2" x 22"
- Door: One swing door, self-closing, right hinged, non-reversible, magnetic sealed gasket, keyed lock
- Shelves: Three shelves (two adjustable/one fixed) with guard rail on back
- Mounting: Low profile roller wheels and leveling legs
- Interior lighting: N/A
- Airflow management: Forced Air technology, patent pending
- External probe access: Rear wall port (1/2") dia.
- Insulation: Cabinet is foamed-in-place with EPA compliant high density urethane foam
- Exterior materials: White powder coated steel
- Access control: Pyxis®, Omnicell® and AcuDose RX® compatible
- General warranty: Two (2) years parts and labor warranty, excluding display probe calibration
- Compressor warranty: Five (5) years compressor warranty
- Product Weight: 100 lbs.
- Shipping Weight: 140 lbs.
- Rated Amperage: 1.74 Amps
- Power Plug/Power Cord: NEMA 5-15 plug, 8 to 10 ft typical, conforms to UL471 requirements, Vaccine storage power cord warning label
- Facility Electrical Requirement: 110-120V AC: 15 A (minimum)
- Agency Listing and Certification: Certified in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. UL, C-UL, ETL, C-ETL listed (either single or dual agency listings) and certified to UL471 standard, hydrocarbon refrigerant safety.
- Included Accessories:
- Pharmacy refrigerator/freezer toolkit and temperature logs
- Digital Data Logger (DDL) meets current CDC requirements for vaccine storage and monitoring. F/C switchable, has a 3-year certification of calibration, and (1) buffered probe in product simulated solution. Min/Max memory along with Alarm event handling. USB port for downloading stored data.
Performance
- Uniformity¹ (Cabinet air): +/- 0.8°C
- Stability² (Cabinet air): +/- 1.2°C
- Maximum temperature variation (Cabinet air): +/- 1.4°C
- Temperature rise after an after 8 sec door openings: Temperature did not exceed 6.4°C at any probe for all required NSF/ANSI 456 testing protocolsᶟ
- Recovery after 3 min door opening: All probes recover to under 8°C within 4.8 min.
- Energy consumption: 1.15 KWh/day⁴
- Average heat rejection: 1.57 KWh/day (224 BTU/h)⁴
- Noise pressure level (dBA): 43 or less installed
- Pull down time to nominal operating temp: 35 min
Note: The CDC requires a data logger or temperature monitoring device to consistently record live temperature readings from the inside of a refrigerator or freezer when storing all types of vaccines. Constant monitoring of vaccines at recommended temperatures is essential for maintaining the efficacy of a vaccine. Failing to do so can lead to wasted vaccines or ineffective vaccines being administered to patients. Please follow the CDC requirements for Vaccine Storage & the COVID-19 Vaccination Program found here.