Enclosure Model: 321
- External Dimensions (Inches): 32.48” W x 24.80” D x 45.67”-48.82” H (min-max)
- Air Flow Version 1P (filter Hepa H14): 256m3/h
- Air Flow version 1C1P (filter Hepa H14 + carbon filter): 164m3/h
- Voltage / Frequency: 90-264V/50-60Hz
- Energetic consumption: 70W
- Side and front panels: Acrylic 8mm
Enclosure Model: 391
- External Dimensions (Inches): 37.37” W x 24.80” D x 45.67”-48.82” H (min-max)
- Air Flow Version 1P (filter Hepa H14): 256m3/h
- Air Flow version 1C1P (filter Hepa H14 + carbon filter): 164m3/h
- Voltage / Frequency: 90-264V/50-60Hz
- Energetic consumption: 70W
- Side and front panels: Acrylic 8mm
Enclosure Model: 483
- External Dimensions (Inches): 50.20” W x 31.50” D x 51.77”-54.92” H (min-max)
- Air Flow Version 1P (filter Hepa H14): 768m3/h
- Air Flow version 1C1P (filter Hepa H14 + carbon filter): 492m3/h
- Voltage / Frequency: 90-264V/50-60Hz
- Energetic consumption: 191W
- Side and front panels: Acrylic 8mm
Enclosure Model: 714
- External Dimensions (Inches): 70.87” W x 31.50” D x 51.77”-54.92” H (min-max)
- Air Flow Version 1P (filter Hepa H14): 1024 m3/h
- Air Flow version 1C1P (filter Hepa H14 + carbon filter): 656m3/h
- Voltage / Frequency: 90-264V/50-60Hz
- Energetic consumption: 261W
- Side and front panels: Acrylic 8mm
Filtration
- Filter HEPA H14: This filtration technology traps particles larger than 0.1 μm with 99.995% efficiency, according to the MPPS method set forth in the EN 1822-1 standard.
- Molecular filter (option): Adding a carbon filter to your vented cabinet allows you to protect its volume from gazes pollutants present in your laboratory environment.
- AS : For organic vapors - BE+ : For organic vapors and acid vapors F : For formaldehyde vapors - K : For ammonia vapors
Equipment:
- Digital control panel: Permanent ventilation control device
- Energy port: To run electrical and fluid lines into the enclosure
- Internal lighting: Compact tubular fluorescent lighting 18W - 500 Lux - IP67