- Volume: 37.43 cu. ft. | 1060L
- Interior Dimensions W x H x D in (mm): 40.94" x 47.24" x 34.46" (1040 x 1200 x 850 mm)
- Exterior Dimensions W x H x D in (mm): 53.94" x 77.56" x 51.18" (1370 x 1970 x 1300 mm)
- Housing: rear zinc-plated steel
- Max. number of internals: 14
- Max. loading of chamber: 440.92 lbs (200 kgs)
- Max. loading per internal: 44.09 lbs (20 kgs)
- Electrical data:
- 115 Volts: 50/60 Hz /approx. 1400 W
- 230 Volts: 50/60 Hz / approx. 1400 W
Temperature Specs:
- Setting temperature range: without light, with humidity: +5 to +70 °C; with light, with humidity: +5 to +70 °C; with light, without humditiy: 0 to +70 °C; without light, without humidity: 0 to +70 °C
- Working temperature range: with light, without or with humidity: 15 to +40 °C; without light, with humidity: +5 (at least 20 below ambient temperature) to +70 °C; without light, without humidity: 0 (at least 20 below ambient temperature) to +70 °C
- Setting accuracy temperature: 0.1°C
- Temperature sensor: 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
Humidity Specs:
- Setting range humidity: 10 – 90 % rh
- Setting accuracy humidity: 0.5%
- Humidity: humidity supply with distilled water from external tank by self-priming pump
- Humidification & Dehumidification: humidification by hot steam generator; dehumidification by cold trap using the Peltier technology
Heating Concept:
- Peltier: energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle)