- Volume: 48.03 cu. ft. | 1360L
- Interior Dimensions W x H x D in (mm): 49.21" x 57.09" x 29.53" (1250 x 1450 x 750 mm)
- Exterior Dimensions W x H x D in (mm): 61.42" x 86.61" x 46.85" (1560 x 2200 x 1190 mm)
- Housing: rear zinc-plated steel
- Max. number of internals: 28
- Max. loading of chamber: 551.16 lbs (250 kgs)
- Max. loading per internal: 66.14 lbs (30 kgs)
- Electrical data:
- 115 Volts: 50/60 Hz /approx. 1720 W
- 230 Volts: 50/60 Hz / approx. 1720 W
Temperature Specs:
- Setting temperature range: without light, with humidity: +5 to +70 °C; with light, with humidity: +5 to +70 °C; with light, without humditiy: 0 to +70 °C; without light, without humidity: 0 to +70 °C
- Working temperature range: with light, without or with humidity: 15 to +40 °C; without light, with humidity: +5 (at least 20 below ambient temperature) to +70 °C; without light, without humidity: 0 (at least 20 below ambient temperature) to +70 °C
- Setting accuracy temperature: 0.1°C
- Temperature sensor: 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
Humidity Specs:
- Setting range humidity: 10 – 90 % rh
- Setting accuracy humidity: 0.5%
- Humidity: humidity supply with distilled water from external tank by self-priming pump
- Humidification & Dehumidification: humidification by hot steam generator; dehumidification by cold trap using the Peltier technology
Heating Concept:
- Peltier: energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle)