- Volume: 26.5 cu. ft. | 749 L
- Interior Dimensions W x H x D in (mm): 48.19" x 76.8" x 32.83" (1224 x 1950 x 834 mm)
- Exterior Dimensions W x H x D in (mm): 40.94" x 47.24" x 23.62" (1040 x 1200 x 600 mm)
- Housing: rear zinc-plated steel
- Installation: on lockable castors
- Max. number of internals: 19
- Max. loading of chamber: 440.9 lbs (200 kgs)
- Max. loading per internal: 66.14 lbs (30 kgs)
- Electrical data: 230 Volts:
- 50 Hz / approx. 1350 W
Temperature Specs:
- Setting temperature range: -10 to +60 °C
- Working temperature range without humidity: -10°C up to +60°C
- Working temperature range with humidity: 10°C up to 60°C
- Setting accuracy temperature: 0.1°C
- Temperature sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Humidity Specs:
- Humidity supply with distilled water from external tank by self-priming pump
- Humidification by hot steam generator
- Active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity - resolution of display 0.1 %, setting accuracy 0.5 %
Heating Concept:
- Cooling: Cooling compressor with climate-friendly refrigerant CO2 (R744), GWP1
- Air jacket: air jacket heating system for gentle all-around heating
- Heating concept: no drying-up of the load caused by the cooling device due to separation of thermal jacket from interior
- Defrosting: highly efficient automatic defrosting system
Control of Standard Components (Light):
- Illumination unit consisting of 4 fluorescent lights with cold white light (standard illuminant D65, 6.500K) and 2 UV-lamps in the spectral range of 320-400 nm
- Daylight and UV-light can be programmed individually via controller
- Cold white light +/- 6 %, for UV-light +/- 10 %