- Performance- ISO 21501-4 unit-to-unit accuracy & reproducibility
- Industry Standards- ISO 21501-4 , ISO 14644-1:2015, FCC, CE
- Light Source- Long Life Laser™ Diode
- Particle Size Analysis Range- 50|10.00|μm / 0.5|10|μm
- Coincidence Loss- 10% at 20,000,000 particles/m3
- Sample Volume- 10.00|µm
- Throughput- 28.32 L per minute
- Zero Count- ISO 21501-4 & JIS B 9921 (1 count or less in 5 min; 95% UCL)
- Count Alarms Display & Cycles- User-defined alarm counts displayed via LED on box or remote LED
- Item Subscriptions Referenced- 2088615-PF-EA
MET ONE 6000P Air Particle Counter
Item#: METONE6000PMET ONE 6000 air particle counters from Beckman Coulter provide a compact design with flexible sizing, flow rate and communication options that address the specific needs of cleanroom environments within the pharmaceutical, semiconductor, hard disk drive and flat panel display industries providing ISO 21501 accuracy for continuous particle monitoring. Reduce initial installation cost with MET ONE 6000P remote air particle counter’s integrated multi-parameter measurements.
Standards
- Built-in flow sensor to assure proper air flow for sampling
- 0.5 µm sensitivity for pharmaceutical and ISO Class 5 clean room monitoring
- Enables compliance with 21 CFR Part 11, FDA cGMP, and EU GMP Annex 1
Operation
- Integrated or external status LED for fast troubleshooting and reduced downtime
- Flow sensor improves operational reliability during continuous monitoring
- Reduced downtime for instrument removal and re-installation for routine calibration/preventative maintenance
Integration
- Multiple communication options for easy FMS integration
- Built-in flow rate sensor controlled vacuum pump for lower installation cost and system complexity
Reliability
- Integrated pump improves reliability across the system – no single point of failure
- Long life laser diode for guaranteed continuous monitoring