- Exterior Dimensions (WxHxD): 31 7/8” x 38 1/8” - 42" x 24 1/4”
- Interior Dimensions (WxHxD): 30 7/8” x 23 3/8” x 19 5/8" - 20 1/8”
- Voltage/Frequency: 100-240 V / 50-60 Hz
- Watts:
- 1P Model: 40
- 1C1P Model: 45
- Decibel Level (dBA):
- 1P Model: 54
- 1C1P Model: 57
- Air face velocity (m/s): 68 fpm
- Air flow (m3/h):
- 1P Model: 117 cfm
- 1C1P Model: 144 cfm
- Side and front panels: 10 mm thick synthetic glass is designed to protect users from harmful UV rays and β (Bêta) emitted
from radioactive isotopes such as: T(3H), 14C, 32P - Structure: Corrosion resistant electro-galvanized steel coated with anti-acid polymer
- Filtration Module: Polypropylene
- Particulate filter (1P): HEPA H14 : This filtration technology traps particles larger than 0.1 μm with 99.995% efficiencyaccording to the MPPS method set forth in the EN 1822-1 standard. ULPA U17 :This filtration technology traps particles larger than 0.1 μm with 99.999995% efficiency according to the MPPS method set forth in the EN 1822-1 standard.
- Carbon filter (optional) (1C): Adding a carbon filter to your enclosure allows protection of your samples from VOCs. AS filter: For organic vapors.
- Particulate pre-filter: Protect particulate filters from dust contaoned in the laboratory environment (only for 1P version)
- Bacterial UV Lights:
- 15W - Wavelength: 254 nm
- 0.08 mJ/ s/cm2
- Internal Lighting:
- LED - IP 44 - 6000k
- 900 lux
- eGuard app (Android or IOS): Mobile app for real time remote control of Smart devices
- Connectivity: RJ45 cable connection to view and change workstation settings (cable included)
- Anemometer: Monitors a drop in pressure that indicates pre-filter or filter replacement is required
- Side panel utility ports: 2 per unit
Optional Accessories (priced separately):
- Benches: Rolling cart (Mobicap) or fixed bench (Benchcap)
- Shelves: Internal metal sliding shelf (only for Benchap)
- Worktop: TRESPA® TOPLABPLUS, Glass or 304L Stainless Steel
- Molecode S: Automatic detection of VOC filter breakthrough