Dimensions:
- Exterior Dimensions (W x D x H): 26.75 x 27.12 x 41.5 in. (679.5 x 689 x 1054 mm)
- Foot Print (W x D): 23.75 x 20.75 in. (603.2 x 527.1 mm)
- Interior Dimensions (W x D x H) 21.5 x 21 x 26.5 in. (546.1 x 533.4 x 673.1 mm)
- Volume: 7.06 ft3 (200 L)
- Net Weight: 235 lbs (106 kg)
Temperature Control
- Control Range: 5° above ambient (to a 30° max ambient) to 55°
- Set-Point Range: 5°C to 55°C (37.0 Default)
- Uniformity: ±0.35°C @ 37°C
- Accuracy: ± 0.1°C
- Recovery: 0.12°C/min. on Average
- Display Resolution: 0.1°C
- Door and Perimeter Heater Control Logic: Base Duty Cycle set to chamber temperature Set-Point Proportional 0-100% [manually adjustable]
- Temperature Sensor Type: Precision Integrated Circuit
CO2 Control Standard:
- Range: 0.1 to 20% 5% Default Set Point
- Accuracy: ±0.1%
- CO2 Recovery: Up to 5% ±0.2% / -0.5 in 5 minutes Average
- CO2 Display Resolution: 0.1%
- CO2 Control Logic: Fixed Algorithm/Manual, Environmental Adaptable
- CO2 Sensor Type: Infrared (NDIR)
Relative Humidity Display and Control
- Control Range: Ambient dependent to 90%
- RH Accuracy: ±3%
- RH Recovery: 90% +5%/-3% within 25 minutes on average
- Display Resolution: 0.1%
- RH Control Logic: Fixed algorithm/manual environmental adaptable
- RH Sensor Type: Precision capacitance bulk-polymer