Performance Specifications |
NU-PR797-400
Nominal 4 foot (1.2m)
|
NU-PR797-600
Nominal 6 foot (1.8m)
|
Product Protection | ISO Class 5 (Laminar Flow) | ISO Class 5 (Laminar Flow) |
Style of Isolator | Bench top/console Isolator | Bench top/console Isolator |
Isolator Construction | Welded stainless steel 16GA, Type 304 pressure tight design | Welded stainless steel 16GA, Type 304 pressure tight design |
Interchange Chamber | ISO Class 5 With Internal/External Sealed Doors | ISO Class 5 With Internal/External Sealed Doors |
Diffuser for Air Supply (Metal) | Non-flammable | Non-flammable |
HEPA Filter Seal Type:
Supply Filter-99.99% Eff. on 0.3 microns Intake Filter-99.99% Eff. on 0.3 microns
|
Neoprene, Spring loaded Under Negative Pressure | Neoprene, Spring loaded Under Negative Pressure |
Standard Services:
Hospital Grade Duplex Outlet
|
One, Backwall | One, Backwall |
Optional Services: Service Valves 3/8" NPT | Up to 3 | Up to 3 |
Work Zone Inches (mm):
Width
Height
Depth (Center of Glove Port)
|
35 1/4 (895)
27 3/8 (695)
23 1/4 (591)
|
59 1/4 (1505)
27 3/8 (695)
23 1/4 (591)
|
Interchange Inches (mm)
Width
Depth (at Work Surface)
Height
|
14 1/8 (359)
24 (610)
27 3/8 (695)
|
14 1/8 (359)
24 (610)
27 3/8 (695)
|
Hinged Viewing Window (Polycarbonate) | Fully closed to fully open | Fully closed to fully open |
Heat Rejected, BTU, Per Hour | 539 | 880 |
Electrical: Volts, AC 60 Hz | UL/UL-C Listed 115 | UL/UL-C Listed 115 |
Amps: Blower/Lights | 3 | 5 |
Amps: Hospital Grade Duplex Amps: Total | 3 | 3 |
12 ft. Hospital Grade Power Cord (one) | 6 | 8 |
14 GA - 3 Wire, 15A | 14 GA - 3 Wire, 15A | |
*Crated Shipping Weight: | 475 lbs./220 kg. | 625 lbs./284 kg. |
*Net Weight: | 425 lbs./193 kg. | 575 lbs./261 kg. |
PharmaGard ES NU-NR800 Recirculating Compounding Aseptic Containment Isolator
Item#: NU-NR800The PharmaGard ES NU-NR800 Compounding Aseptic Containment Isolator from NuAire uses a Restricted Access Barrier Systems (RABS) which creates a negative pressure ISO Class 5 controlled environment for the aseptic compounding of hazardous drugs and complies with USP <800> requirements. These isolators provide multiple levels of protection in order to eliminate contamination and minimize the possibility of chemotherapy agents or other hazardous materials from entering or exiting the work area.
- Locking interchange compartment
- Incorporates the latest advancements in laboratory ergonomics
- Quick change, air-tight waste disposal system
- Ultra-high efficiency motor increased filter life and lowers energy costs
- Stable, consistent unidirectional airflow
- HEPEX Zero Leak Airflow System
- Technology to reduce unit noise and vibration