SPECTROstar Omega
-
Detection Modes
UV/vis absorbance spectra
-
Measurement Modes
Endpoint and kinetic
Spectral scanning (absorbance)
Well scanning -
Microplate Formats
6 to 1536-well plates, user-definable
LVis Plate with 16 low volume microspots (2 μL) -
Microplate Carrier
Robot compatible
-
Light Sources
High energy xenon flash lamp
-
Detectors
CCD Spectrometer
-
Wavelength Selection
UV/Vis Absorbance Spectrometer
Full spectra or 8 distinct wavelengths in < 1 sec / well -
Spectral Range
Spectrometer
220 - 1000 nm -
Sensitivity
Full spectrum captured in < 1 s/well
Selectable spectral resolution: 1 nm, 2 nm, 5 nm, 10 nm
OD range: 0 to 4 OD
Accuracy: < 1% at 2 OD
Precision: < 0.5% at 1 OD and < 0.8% at 2 OD
Reproducibility: ±0.005 OD for 0 - 2 OD range -
Read Times
Full spectrum from 220 to 1000 nm in less than 1 sec/well
-
Reagent Injection
Up to 2 built-in reagent injectors
Injection at measurement position (6 to 384-well)
Individual injection volumes for each well (3 to 500 µL)
Variable injection speed up to 420 µL / s
Up to four injection events per well
Reagent back flushing -
Shaking
Linear, orbital, and double-orbital with user-definable time and speed
-
Purge Gas Vent
System to inject an atmosphere or to pull a vacuum into the reader
-
Incubation
+4 °C above ambient up to 45 °C or 65 °C
The upper heating plate of the incubation chamber operates at 0.5 °C more than the lower plate. This prevents condensation build-up on the lid or sealer.
-
Software
Multi-user Reader Control and MARS Data Analysis Software included
FDA 21 CFR Part 11 compliant -
Dimensions
Width: 44 cm, depth: 48 cm, height: 30 cm; weight: 28 kg
-
Optional Accessories
-
Atmospheric Control Unit (ACU)
Actively regulates O2 and CO2 - 0.1-20%
-
LVis Plate
Sample capacity: sixteen separate microdrop wells for 2 µL samples; one standard cuvette position for up to 1 mL samples.
Quality control internal standards (optional): four NIST-traceable optical density filters (approximate values of 0.1, 0.3, 0.6 and 1.0 OD); one holmium oxide filter for wavelength accuracy
Dimensions: conforms to SBS standards for microplates. -
Stacker
Magazines for up to 50 plates - continuous loading feature
-
THERMOstar
Microplate incubator and shaker
-
Upgrades
Please contact your local representative for upgrades including options such as detection modes, reagent injectors, etc.